EHDS and the Cyber Resilience Act: what applies to health software

4 min read

The European Health Data Space (EHDS, Regulation (EU) 2025/327) turns software for electronic health records into a CE-marked product. Many of these systems also fall under the Cyber Resilience Act – and the EHDS Regulation amended it specifically for this. This guide shows what applies today and what is added from 2029.

What an EHR system is

Under the EHDS Regulation, an EHR system is any system whose software – alone or together with hardware – stores, intermediates, exports, imports, converts, edits or views personal electronic health data in the priority categories, and which the manufacturer intends to be used by healthcare providers for patient care or by patients to access their data.

  • Priority categories: patient summaries, electronic prescriptions, electronic dispensations, medical imaging studies with related reports, medical test results (such as laboratory results) and discharge reports.
  • General purpose software used in a healthcare environment is excluded.
  • An EHR system offered as a service is also deemed to be put into service – so the EHDS obligations also cover cloud offerings.

Source: Art. 2 (definition of ‘EHR system’) EHDS Regulation · Art. 14(1) EHDS Regulation · Art. 25(2) EHDS Regulation · Art. 26(2) EHDS Regulation

Already today: the Cyber Resilience Act

If an EHR system is delivered as software – for example installed practice or hospital software – it is a product with digital elements. The Cyber Resilience Act then already applies: the reporting obligation for actively exploited vulnerabilities and severe security incidents since 11 September 2026, all other obligations from 11 December 2027.

Pure cloud services designed and developed outside the responsibility of a manufacturer of a product with digital elements do not fall under the Cyber Resilience Act; cloud services such as software as a service may be covered by the NIS 2 Directive. Medical devices and in vitro diagnostic medical devices are excluded from the Cyber Resilience Act – the medical device regulations apply to them.

Wearable products for health monitoring that are not medical devices are expressly listed in the Cyber Resilience Act as important products of class I.

Source: Art. 14 CRA · Art. 71 CRA · Recital 12 CRA · Art. 2(2) CRA · Annex III class I point 19 CRA

Obligations under the EHDS

EHR systems must include two harmonised software components: the European interoperability software component and the European logging software component. The essential requirements for these components are set out in Annex II.

  • Draw up the technical documentation before placing the system on the market and keep it up to date.
  • Provide the information sheet and instructions for use free of charge.
  • Draw up the EU declaration of conformity and affix the CE marking; if other EU law requiring its own declaration applies to the system, a single declaration covering all legal acts is drawn up.
  • Before placing on the market, test the harmonised components in the digital testing environments of the Commission or the Member States.
  • Register the system in the public EU database for EHR systems.
  • In case of non-conformity, take corrective action without undue delay, recall or withdraw the system and inform the authorities.

Source: Art. 25(1) EHDS Regulation · Annex II EHDS Regulation · Art. 30(1) EHDS Regulation · Art. 39 EHDS Regulation · Art. 40 EHDS Regulation · Art. 49 EHDS Regulation

From when

  • The regulation applies in principle from 26 March 2027.
  • For EHR systems intended to process patient summaries, electronic prescriptions or dispensations, the rules on the harmonised components apply from 26 March 2029.
  • For imaging, test results and discharge reports from 26 March 2031.
  • For EHR systems manufactured and used within healthcare institutions, Chapter III applies from 26 March 2031.

The Commission adopts common specifications for the essential requirements by 26 March 2027 by means of implementing acts. Only then will it be clear in detail what the components must do technically.

Source: Art. 105 EHDS Regulation · Art. 36(1) EHDS Regulation

One documentation for the Cyber Resilience Act and the EHDS

The EHDS Regulation amended the Cyber Resilience Act. Manufacturers of products with digital elements classified as EHR systems demonstrate conformity with the essential cybersecurity requirements of Annex I CRA by applying the conformity assessment procedure under Chapter III of the EHDS Regulation (new Art. 32(5a) CRA).

A single set of technical documentation is drawn up containing the information under Annex VII CRA and that required by the EHDS Regulation (Art. 31(3) CRA); the cybersecurity risk assessment may be part of the risk assessment required there (Art. 13(4) CRA). Keeping the documentation under the Cyber Resilience Act in good order today lays the foundation for the EHDS evidence.

Source: Art. 104 (amendment of the Cyber Resilience Act) EHDS Regulation · Art. 31 CRA · Art. 32 CRA · Art. 13 CRA · Annex VII CRA

Medical devices connected to EHR systems

Manufacturers of medical devices or in vitro diagnostic medical devices who claim interoperability with the harmonised components of EHR systems must meet and demonstrate the requirements for these components in Annex II, Section 2.

Source: Art. 27(1) EHDS Regulation · Annex II Section 2 EHDS Regulation

In practice

  • Clarify which products are EHR systems, which data categories they process and whether they are delivered or offered as a service – this determines the obligations under the Cyber Resilience Act and the EHDS deadline.
  • For delivered software, implement the obligations under the Cyber Resilience Act now: reporting process under Art. 14, SBOM, vulnerability handling, support period.
  • Structure the technical documentation so that the EHDS information can be added later instead of keeping a second documentation.
  • Follow the Commission's common specifications (by 26 March 2027) and plan the components afterwards.